PERRY JOHNSON LABORATORY
ACCREDITATION, INC.
Certificate of Accreditation
Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of:
Nagano Science Co., Ltd. Calibration Laboratory
1-10 Amashimmachi Takatsuki, Osaka 569-1106
(Hereinafter called the Organization) and hereby declares that Organization is accredited
in accordance with the recognized International Standard:
ISO/IEC 17025:2005
This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory quality management system
(as outlined by the joint ISO-ILAC-IAF Communiqué dated January 2009):
Calibration of capacitance-type relative humidity sensor and digital thermometer
(As detailed in the supplement)
Accreditation claims for such testing and/or calibration services shall only be made from addresses referenced within this certificate. This Accreditation is granted subject to the system rules governing the Accreditation referred to above, and the Organization hereby covenants with the Accreditation body’s duty to observe and comply with the said rules.
Initial Accreditation Date: Issue Date: Expiration Date: October 28, 2013 August 3, 2017 September 30, 2019 Accreditation No.: Certificate No.:
72624 L17-337
The validity of this certificate is maintained through ongoing assessments based on a continuous accreditation cycle. The validity of this certificate should be
confirmed through the PJLA website: www.pjlabs.com For PJLA:
Tracy Szerszen
President/Operations Manager Perry Johnson Laboratory Accreditation, Inc. (PJLA) 755 W. Big Beaver, Suite 1325
Certificate of Accreditation: Supplement
Nagano Science Co., Ltd. Calibration Laboratory
1-10 Amashimmachi Takatsuki, Osaka 569-1106
Mihoko Kondo Phone: 072-681-8800
Accreditation is granted to the facility to perform the following calibrations:
Issue: 08/2017 This supplement is in conjunction with certificate #L17-337 Page 2 of 4
Thermodynamic
F MEASURED INSTRUMENT, QUANTITY OR GAUGE RANGE OR NOMINAL DEVICE SIZE AS APPROPRIATE CALIBRATION AND MEASUREMENT CAPABILITY EXPRESSED AS AN UNCERTAINTY (±) CALIBRATION EQUIPMENT ANDREFERENCE STANDARDS USED
Capacitance-type relative humidity sensor Rotronic temperature and humidity probe 11% RH to 80% RH 25 °C to 40 °C
1.2 %RH Temperature humidity generator (Michell Inst.: Opti-Cal)
Chilled mirror dew point hygrometer (Michell Inst. : S8000 Remote) Dew point: -9 °C to 37 °C Temperature data logger
(Vaisala: 1000-22N, 1000-21N, VL-1400-44N)
Temperature: -25 °C to 70 °C
Humidity Calibration Procedures for Rotronic Temperature and Humidity Probe
(by Calibration Item List) (SZY0002) based on JIS B 7920 Capacitance-type relative humidity sensor Testo multi environmental measurement system 11 %RH to 90% RH@ 25 ℃
1.4 %RH Temperature humidity generator (Michell Inst.: Opti-Cal)
Chilled mirror dew point hygrometer (Michell Inst. : S8000 Remote) Dew point: -9 °C to 37 °C Temperature data logger
(Vaisala: 1000-22N, 1000-21N, VL-1400-44N)
Temperature: -25 °C to 70 °C
Humidity Calibration Procedures for Testo Multi Environmental Measurement System (by Calibration Item List) (SZY0005) based on JIS B 7920 11 %RH to 80 %RH@ 40 ℃ 1.1 %RH Digital thermometer Rotronic temperature and humidity probe
-25 °C to 0 °C 0.15 °C Temperature and humidity chamber (Nagano Science Co., Ltd.: LH43-13P) Temperature: -25 °C to 65 °C
Temperature data logger
(Vaisala: 1000-22N, 1000-21N, VL-1400-44N)
Temperature: -25 °C to 70 °C
Temperature Calibration Procedures for Rotronic Temperature and Humidity Probe (by Calibration Item List)(SZY0001)
Certificate of Accreditation: Supplement
Nagano Science Co., Ltd. Calibration Laboratory
1-10 Amashimmachi Takatsuki, Osaka 569-1106
Mihoko Kondo Phone: 072-655-9848
Accreditation is granted to the facility to perform the following calibrations:
Thermodynamic
F MEASURED INSTRUMENT, QUANTITY OR GAUGE RANGE OR NOMINAL DEVICE SIZE AS APPROPRIATE CALIBRATION AND MEASUREMENT CAPABILITY EXPRESSED AS AN UNCERTAINTY (±) CALIBRATION EQUIPMENT ANDREFERENCE STANDARDS USED
Digital thermometer Rotronic temperature and humidity probe
0 °C to 65 °C 0.15 °C Temperature and humidity chamber (Nagano Science Co., Ltd.: LH43-13P) Temperature: -25 °C to 65 °C
Temperature data logger
(Vaisala: 1000-22N, 1000-21N, VL-1400-44N)
Temperature: -25 °C to 70 °C
Temperature Calibration Procedures for Rotronic Temperature and Humidity Probe (by Calibration Item List)(SZY0001) Digital thermometer Testo multi environmental measurement system temperature probe -20 °C to 0 °C 0.15 °C Temperature calibrator (AMETEK: RTC-158A) Temperature data logger
(Vaisala: 1000-22N, 1000-21N, VL-1400-44N)
Temperature: -25 °C to 70 °C
Temperature Calibration Procedures for Testo Multi Environmental Measurement System (by Calibration Item List) (SZY0004) 0 °C to 100 °C 0.14 °C Temperature calibrator
(AMETEK: RTC-158A) Temperature data logger
(Vaisala: 1000-22N, 1000-21N, VL-1400-44N)
Temperature: -25 °C to 70 °C Thermistor thermometer
(Fluke:1504 Indicator, 5610-6 Sensor) Temperature: 0 °C to 100 °C
Temperature Calibration Procedures for Testo Multi Environmental Measurement System (by Calibration Item List) (SZY0004)
Certificate of Accreditation: Supplement
Nagano Science Co., Ltd. Calibration Laboratory
1-10 Amashimmachi Takatsuki, Osaka 569-1106
Mihoko Kondo Phone: 072-655-9848
Accreditation is granted to the facility to perform the following calibrations:
Issue: 08/2017 This supplement is in conjunction with certificate #L17-337 Page 4 of 4
Thermodynamic
F MEASURED INSTRUMENT, QUANTITY OR GAUGE RANGE OR NOMINAL DEVICE SIZE AS APPROPRIATE CALIBRATION AND MEASUREMENT CAPABILITY EXPRESSED AS AN UNCERTAINTY (±) CALIBRATION EQUIPMENT ANDREFERENCE STANDARDS USED
Digital thermometer Testo multi environmental measurement system temperature and humidity probe 0 °C to 60 °C 0.2 °C Temperature calibrator (AMETEK: RTC-158A) Temperature data logger
(Vaisala: 1000-22N, 1000-21N, VL-1400-44N)
Temperature: -25 °C to 70 °C
Temperature Calibration Procedures for Testo Multi Environmental Measurement System (by Calibration Item List) (SZY0004) 1. The CMC (Calibration and Measurement Capability) stated for calibrations included on this scope of accreditation represent
the smallest measurement uncertainties attainable by the laboratory when performing a more or less routine calibration of a nearly ideal device under nearly ideal conditions. It is expressed at a confidence level of 95 % using a coverage factor k (usually equal to 2). The actual measurement uncertainty associated with a specific calibration performed by the laboratory will typically be larger than the CMC for the same calibration since capability and performance of the device being calibrated and the conditions related to the calibration may reasonably be expected to deviate from ideal to some degree. 2. The laboratories range of calibration capability for all disciplines for which they are accredited is the interval from the
smallest calibrated standard to the largest calibrated standard used in performing the calibration. The low end of this range must be an attainable value for which the laboratory has or has access to the standard referenced. Verification of an indicated value of zero in the absence of a standard is common practice in the procedure for many calibrations but by its definition it does not constitute calibration of zero capacity.
3. The presence of a superscript F means that the laboratory performs calibration of the indicated parameter at its fixed location. Example: Outside MicrometerFwould mean that the laboratory performs this calibration at its fixed location.
ペリージョンソン ラボラトリー
アクレディテーション インク
認 定 証
ペリージョンソン ラボラトリー アクレディテーション インクは、
下記の試験所を審査しました。
ナガノサイエンス株式会社 校正室
〒569-1106 大阪府高槻市安満新町 1-10
ここに本組織が、以下の認知された国際規格に基づき、認定されたことを証します。ISO/IEC 17025:2005
本認定により、以下の範囲及び試験所品質マネジメントシステムの運営における技術的能力を 実証するものとします。(2009 年 1 月発行 ISO-ILAC-IAF 共同コミュニケに準ずる)静電容量式相対湿度計及びデジタル温度計の校正
(詳細は付属書に記述)
上記試験及び/又は校正サービスに対する認定資格は本認定証内で言及された住所のみを対象とする。本認定は、 上記規格の認定を管理するシステム規定に従い授与され、組織はその規定を遵守し、認定機関の任務を尊重する ことをここに誓約する。 尚、本認定証は日本語翻訳版であり、英文の認定証を正式のものとする。 初回認定日 発行日 認定証有効期限 2013 年 10 月 28 日 2017 年 8 月 3 日 2019 年 9 月 30 日 認定番号 認定証番号 72624 L17-337 この認定証の有効性は、持続された認定に基づく継続審査を通して維持されています。 PJLA ウェブサイト (www.pjlabs.com)でご確認いただけます。 尚、本認定証は日本語翻訳版であり、英文の認定証を正式のものとする。 PJLA___________________
トレーシー サーツェン プレジデント/オペレーションマネージャーPerry Johnson Laboratory Accreditation, Inc. (PJLA)
755 W. Big Beaver Rd., Suite 1325 Troy, Michigan 48084
認 定 証 付 属 書
2017 年 8 月発行 本付属書は、認定証番号 L17-337 の認定証を補足するものである。 2/4 頁ナガノサイエンス株式会社 校正室
〒569-1106 大阪府高槻市安満新町 1-10
近藤 美穂子 Tel: 072-681-8800 本認定を、上記組織の実施する下記試験について授与する。 温度的校正F 校正を受けた機器, 計量または計測器 範囲または必要に応じて 基準装置サイズ 不確かさとして 表現された 校正測定能力(+/-) 使用される校正機器 および参照標準 静電容量式 相対湿度計 Rotronic 製温湿度 プローブ 11%RH~80%RH@ 25 ℃~40 ℃ 1.2 %RH 温湿度発生器 (Michell Inst.製:Opti-Cal) 鏡面冷却式露点計(Michell Inst.製:S8000 Remote) 露点:-9 ℃~37 ℃ 温度データロガー (Vaisala 製: VL-1000-22N,VL-1000-21N,VL-1400-44N) 温度:-25 ℃~70 ℃ Rotronic 製温湿度プローブ湿度校正手順書 (校正品目リストによる)(SZY0002) JIS B 7920 に基づく 静電容量式 相対湿度計 Testo 製マルチ環境 計測システム 11 %RH~90 %RH@ 25 ℃ 1.4 %RH 温湿度発生器 (Michell Inst.製:Opti-Cal) 鏡面冷却式露点計
(Michell Inst.製:S8000 Remote) 露点:-9 ℃~37 ℃ 温度データロガー (Vaisala 製: VL-1000-22N,VL-1000-21N,VL-1400-44N) 温度:-25 ℃~70 ℃ Testo 製マルチ環境計測システム 湿度校正手順書(校正品目リストによる) (SZY0005) <JIS B 7920 に基づく> 11 %RH~80 %RH@ 40 ℃ 1.1 %RH デジタル温度計 Rotronic 製温湿度 プローブ -25 ℃~0 ℃ 0.15 ℃ 恒温恒湿槽 (ナガノサイエンス製:LH43-13P) 温度:-25 ℃~65 ℃ 温度データロガー (Vaisala 製: VL-1000-22N,VL-1000-21N,VL-1400-44N) 温度:-25 ℃~70 ℃ Rotronic 製温湿度プローブ温度校正手順書 (校正品目リストによる)(SZY0001)