(平面白〕
,
‑
, p o
nU
唱・a
一 一 一 一 C N S ‑ O N
C N S ‑ O F F
mm,mõ勺::~::::30
p︑d
n u
司A
E r ‑ ‑
ト ト ト ト
n u n u
今4A HZ
・
0‑・
4 )凶ZU出2M﹀a
z‑
4国
寸
i =
104f f i
トー 103
トーー
:z:: 2ト ロ 10
" "
トー
雲
101トー
10(AI
ーーー回ー・
2 4 6
N E U T R O N W A V E L E N G T H
図 ‑ 3 . N E U T R O N G A I N B Y K U R C N S I
Llc . H 2 )
。
占‑‑‑l..2 ̲ ̲ 4 ' 6N E U T R O N W A V E L E N G T H
8o o B
101A)図ー2 .
IlE U T R O H I I I T E N S I T I E S
1MC N S (
lIC,H 2) O N A N D O F F
‑39 ‑
Guiding Field for The Neutron Life Measurement
Masayoshi Wake and Keisuke Maehata
KEK National Laboratory for High Energy Physics
Introduction
The life time of neutron can be measured by counting electrons or protons produced by the decay of neutrons in neutron beam n i t h Known linear density. The major problem in this measurement is the definition of the decay volume and the accurate counting of decayed particle which is not too large compared to the background signals. The low energy electron makes small circular orbit around the flux line and. as a result, travels along the flux line. Therefore one can gather electrons in one direction by magnetic field. By this, counting rate of the decayed electron for the same dimension of the detector, which means the same back ground, can be significantly increased. If the guiding field line crosses the beam and the decayed electron follows the field line the decay volume can be defined by the field line. The ambiguity of the decay volume become less than the diameter of the electron orbit. The guiding field in the neutron life experiment is effective both for the definition of the decay volume and the counting efficiency.
Electron Trapping by The Inhomogeneous Field
If the electrons simply follow the field line, the most efficient guiding field would be a long solenoid coil because all the electrons produced in the neutron beam line are gathered at the both ends of the solenoid.
Fig. 1 shows the magnetic field in such an arrangement. The electron counters can be small and h'ince the back ground for this measurement is very small. This type of magnet was used in the experiment in ref. 1. The problem in this type of magnet is the definition of the decay volume.
Since the decay volume is defined by the curved field at both ends, the magnet has to be very Ion? to have accuracy In the decay volume. If the magnetic field is perpendicular to the beam line. the definition of the decay volume become much easyer. Although. the gathering efficiency of the decayed electron can not be so large as a long solenoid. Even if the magnetic field is perpendicular to the beam line. there is another confusion in the decay volume. Low energy electrons are trapped in a bulged field. This is a well known effect applied for mirror fusion facilities. As a matter of fact, more than 10% of the decayed electron was lost in the usual magnetic field between pole pieces which has slight
Inevitable bulge of the field. (ref2)
Superconducting Split Solenoid
The magnet for the guidir.7 field of the neutron life measurement should not have buldging field. It is very difficult to reduce the buldge of the
field for a conventional magnet to the order of less than 10~ . But if the magnet is composed of a pair of split superconducting solenoid coil, the magnetic field in the experimental area is rather spreded at the outside of the coil which means there is no buldging in the field at all.
. , .
o
Guiding Field for The Neutron Life Measure田ent
lIasayoshi胃ake and Kel5uke Maehata
KEK National Laboratory for High Energ吉PhYsics
Introduction
官he llfe tl皿e of neutron can be mea5ured by countlng electron5 or proton5 produced by the decay of neutrons 1n neutron bea皿withKnawn linear densit~ The major problem 1n th1s皿easurement 15 the definltion
。
f the decay volu皿e and色he accurate counti~ of decayed partic1e wb1cb 19 not too 1arge compared to the background signals. Th. low energy e1ectron血akes small clrcular orbit around the flux l1ne and. as a result.色ravels a1005 the flux 11ne. Therefore one ca且gather e1ectro且s ln one dlrectlon by mag冨etlc fleld. By this. cou且tlngrate of the decayed electron for the same dlmenslo且 。f the detector. whlch皿eans the same back grou且d. ca且be slgnlficantly increased. If the guldlng fleld l1ne crosses the bea皿 and
色he decayed electron fol1oW5 the fleld line the decay votu皿.e can be deflned by the fleld l1ne. The a皿bigultyof the decay volume becollle le5s tban the dlameter of tbe electron orbit. The guldlO5 field ln the oeutron life experiment Is effectlve botb for the deflnltlon oC the decay volttne and the counting efficiency.
Electron Trapplng by The Ir,hOmogeneous Fleld
1f the eleotrons simply follow the fleld 1Ioe. the回ost errlclent guldlng field would bc a long soleooid ooil because all thc electroo r.produced in the neutron bea皿 line are gathered at thc b。色h ends of the soleoold.
Fig.l shows the magnetic field in such sn arrnnge田.ot. The electron cou且ters can be smal1 and h'Jnce the back ground for this皿easurementis very s田al1. Th15 type of m~gnet was used ln the experlment 10 ref. 1. The proble田 10 this type of mnonet is the defioltion of the decay volu血e. Sloce the decay volume 1. d号fioed by the curved field at both ends. the magnet haa to be very 100耳 to have accuracy 10 the decay volume. If the
皿agnetic field 18 perpcod.c111ar to the beam 110e. the deflnltlon of the decay volu皿e become mucb edsyer. Althoug~ the gatbering efflcleocy of the decayed electron can not be 80 large as a long Bolenold. Even if色he magnetic Cleld 15 perpendicular to the bea皿 llne. there Is ano色her confuslon 1n the decay volume. Low energy electrons are trapped ln a bulged f leld. Thls 15 a well 孟nown effect applled for mirror !u810n fac111tles. As a matter of fact, more than 10兎 of thc decayed clectron was lost in the usual皿agr.etlc Cleld between pole pieces which has s11ght
lnevitable bulge of the field. (refZ)
Superconducting Spl1t Soleロoid
The lIIagnet for the guidlr." field oC the neutron J lf" 血easure~ent should not have buldging Cleld. 1t Is very dlcricult to reduce the bu1dge of the
field for a conventional magnet to the order of less than 10・ B uttf the magnet is composed of a palr oC spl1t Buperconducting 801eoold coll. the皿agneticf1e Id 且 the exPerlmental area 18 rather spreded at the outslde of the col1
・
hichmeans there 18 no bUldglng ln the fleld at all.The flux line of the designed magnet is shown in Fig. 2. The central field of the magnet is 1. 5 T. The inner diameter of the coil is 500 mm and the gap between two solenoid coils is 80 mm. The operation of the magnet is made with the current of ISO A. The stored energy and the inductance of the magnet are 162 kJ and 14. 4 H. respectively. The conductor of the magnet is planed to have 1. 5 mm diameter pure copper matrix with 50 % of its cross section filled by 46. 5%Nb-Ti fine filaments.
The coil has 2559 turns in each block of the split solenoid winding.
Magnet Construction
The real construction and assemble of the magnet has to be designed after a careful consideration of liquid helium consumption. Figure 3 is the cross sectional drawing of the magnet. The support structure of the coil is the inner vessel of the cryostat itself. the coil is mounted in the inner vessel of the cryostat by shrink fitting. The mechanical support of the inner vessel in the cryostat can be made at the window for tit beam line. A kind of bellow structure can keep the inner vessel and the coil at the center allowing the thermal contraction in the radial direction.
The field map of this coil is shown in Fig. 4. The largest magnetic field on the conductor exceeds 3 T even the central field of the magnet is as low field as 1. 5 T. Since the operation current of this magnet is 150 A, it is possible to hold the magnetic field in persistent current mode with a superconducting switch. By the elimination of the excess heat load, the helium consumption of the cryostat may be minimized down to a hundred litters per week. The protection of the magnet against a quench should be made by the protection resistance in the cryostat. The size of the protection resistance can be rather small because the large gas flow due to the evaporation of the liquid helium will cool the resistance if the
resistance is located in the proper position. The main difficulty of the protection of the magnet is in the normal resistance of the superconducting switch. The protection resistance of 3 0 requires at least 300. off resistance in the superconducting switch which is very large compared to the standard size. Figure S is the simulated quench result assuming large enough switch resistance. In this case, the magnet can survive through quenches with its maximum temperature of 44 K.
Conclusion
the neutron life measurement can be made with the aid of superconducting magnet field. The magnet which produces a good field can be constructed by use of present superconducting magnet technology. The preliminary design of the magnet was shown in this report.
References
1. P. Bopp et. al.. Phys. Rev. Lett. 56(1986)919 Z. C J. Christensen et. al.. Phys. Rev. D5( 1972] 1628
" " "
h岨
The flux line of tbe designed magnet Is SbOWO ln Flg. 2. Tbe ceotral fleld of the皿agnet Is 1. 5 T. The ln且er diameter of tbe col1 15 500皿皿
aOd the gap between two solenold coi18 18 80 ~ The operation of the mag百et i9 made with the current of 150 ~ The stored energy and the induclance of the magnel are 162 kJ and 1~4 払 respectlvelY. The conductor of tbe血ag百et 18 planed to have 1. 5 m田diameterpure copper matrix with 50包 。f lt8 crQSS sectloo fll1ed by 46. 5%Nb‑Ti fine fila皿ents. The col1 has 2559 turos 10 each block of tbe spl1t solenoid曹inding.
Uag百et Constructio且
The real constructl。且 a且d assemble of the magnet has to be deslgned after a careful consideratlon of liquld hellu血 con5umptio~ Flgure 3 15 the
C~oss sectlonal drawlog of the血agnet. The support s.ruc色ure of the 0011 18 the lnner vessel of the cr~ostat ltself. the coil 18 mouoted in the lnner vessel of tbe oryostat by sh.rlnk fltting. The mechanlcal support of the ln且ervessel 10 the cryostat can be made at the window for t~~ beam l1ne. A klnd of bel10w structure ca且keep tbe lnner vessel and tbe col1 at the ceoter al10wing the thermal contraction ln tbe radlal directlo~
The field皿apof thls coil 18 shown ln Fl~ ~ The largest magnetlc field
。
n the conductor exceedsa
T even the central field of tbe皿agnet ls as low fleld as 1. 5 ~ Slnce the operatioo curreot of this血agnet Is 150 A it Is possible to hold the magnetlc field in persistent currcnt mode wlth a superconductlog s曹itch. By the el1田lnatlonof tbe excess heat load. the helium consumptio且 。f the cr~ostat ma~ be miol田lzed down to a bundred 11tters per wee~ The protectlon of the magnet agalnst a quench should be made by the protection resistance 且 tbe cryostat. The slze of tbe protectlon resistance can be rather small because the large gas flow due to tbe evaporatlon of tbe liquid heliu血 wl11 cool the resistance lf theresistance is located In ¥he proper posltlo~τhe malo dl!flculty of the protectlon o( tbe magnet 18 10 the oormal res18tence ot the superconductilllf switch. The protectlon resistance of 3D requires at least 80D. off resl8ta!¥ce in the superconduc色Ing switch whlch Is very large compared to the sta且dard 81ze. Figure 5 ls the 81皿ulatedquench resul t assum11llf large eoough swltch res 1staoce. 10 this case. the皿agnet can survive through queoches wltb its田axlmu皿 temperature of 44 K
Conclusion
the neu色ron l1fe me&surewent can be 皿ade with the ald of supercondu~tlng
magnet fleld. The田agnet which produces a good fleld can be coostructed by use of preseot 5upercooductJog皿agoet tecbnology. The prelimlnary deslgn of tbe magnet .as sho
・
o ln this report.Refere且ces
1. P. Bopp et. al.. Ph1s. Rev. Lett. 56! 1986)919 2. C. J.
ロ
lrIste且seoet. al.. Phys. Ruv. D5{ 197211628IN3
Fig-. 1 Long Solenoid Coil Arrangement for Neutron Life Measurement The flux line at both end of the solenoid can be bent to the radial direotion by placing compensation ooils.
Fig. 2 Flux Lines In The Split Solenoid
The flux lines are actually the equal vector potential lines calculated by POISSON group programs.
Fig. 3 Split Solenoid Magnet for Neutron Life Measurement
The coil is assembled in the oryostat prior to the welding of the cryostat. The dimensions are subject to chage for the reduction of heat load.
o BH01
cb 7 o C J . O O ' « . 0 0 ' C 00 £ 0 0 ,Z Cy 12.00 U 0 0 : 6 . ( < 0 I B tit 2C.C0 <*.,00 ^'t 00
Fig. 4 Field Map of The Magnet Lines are at every 0. 05 tesla.
01>.
~
Fig. 1 Long So1eno1d Co1l Arrangement for Neutron u1fe Measurement The flux 11ne at both end of the solenoid can be bent to the radial dire。色lonby placing compensation coils.
Fig. 2 Flux Lines 10 The Split Solenold
The flux lines are actually the equal vector potent1al llnes calculated by FOISSON group progra皿s.
Flg. a Split Solenoid Magoet for Neutron Llfe Measure皿ent
The col1 Is assembled in the cryostat prlor to the weldlng of the cryostat. The dlmenslons arの sUbject to cbage fOr the reduct10n of heat load.
fluOI
Fig. 4 Field Map of The Magnet Llnes are at every ~05 tesl~